检验医学与临床Issue(z2):41-42,43,3.DOI:10.3969/j.issn.1672-9455.2015.26.017
2007~2014年参加全国凝血试验室间质评结果分析
Analysis of the results of the national coagulation test room in 2007-2014
马升俊 1王晶 1许成良1
作者信息
- 1. 广西医科大学第三附属医院检验科,广西南宁 530031
- 折叠
摘要
Abstract
Objective Through the retrospective analysis of 2007 - 2014 in the coagulation test EQA (EQA) results ,summarized and analyzed the reason of losing control in the EQA and control experience ,in order to improve the inspection quality .Methods Statistical analysis of the results of the quality assessment of the chamber was con‐ducted by using the ability of comparative test (PT ) and bias .Results The coagulation rate is 1 .25% PT project out of control ,the INR control rate was 2 .50% ,APTT and Fbg control rate was 0% ;all PT scores were more than 80% ,summary of results are satisfactory ;in addition to control data ,the project quality assessment results of each bi‐as were less than the standard 1/2 evaluation .Conclusion To strengthen personnel training ,improve the level of in‐spection personnel ;good equipment maintenance and maintenance ,timely understanding of the operation state of the instrument and reagent quality ,timely on the external quality assessment results were analyzed ,to ensure the inspec‐tion quality .关键词
凝血试验/室间质评/能力对比试验/偏倚Key words
blood coagulation test/external quality assessment (EQA )/ability of comparative tests (PT )/bias引用本文复制引用
马升俊,王晶,许成良..2007~2014年参加全国凝血试验室间质评结果分析[J].检验医学与临床,2015,(z2):41-42,43,3.