检验医学与临床Issue(1):29-30,2.DOI:10.3969/j.issn.1672-9455.2016.01.012
应用 CantoⅡ流式细胞仪建立14~16岁人群 CD64正常参考值的探讨∗
Establishment of normal reference value for CD64 in healthy adolescents aged 14 - 1 6 years by Canto Ⅱ flow ;cytometer
摘要
Abstract
Objective To investigate the normal reference value establishment of CD64 in the healthy adoles-cents aged 14-1 6 years by the CantoⅡ flow cytometer.Methods The peripheral blood samples collected from 52 healthy adolescents aged 14-1 6 years were detected the CD64 expression by using the BD CantoⅡ flow cytometer, and the results were statistically analyzed.Results The expression of CD64 was highest in monocytes,but almost no expression was found in lymphocytes.The CD64 mean fluorescence intensity (MFI)in neutrophils was 782.78,its 95% normal reference range was 21 1.78 - 1 353.78,the 95% normal reference range of CD64 index was less than 1.53.Conclusion The method for detecting the normal reference value of CD64 in the healthy adolescents by the CantoⅡ flow cytometer is preliminarily established,providing a reference for the basic researches and clinical diano-sis.关键词
CantoⅡ流式细胞仪/CD64/正常参考值Key words
CantoⅡ flow cytometer/CD64/normal reference value引用本文复制引用