中国医疗设备Issue(1):126-127,148,3.DOI:10.3969/j.issn.1674-1633.2016.01.039
我院呼吸机的质量控制
Quality Control of Ventilators in Our Hospital
万胜春 1何敏1
作者信息
- 1. 解放军第105医院 设备科,安徽 合肥230031
- 折叠
摘要
Abstract
Objective The quality of ventilators in hospital is checked regularly, in order to guarantee the safety of clinical.Methods According to the test items and methods for ventilators from“Standard for Quality Control of Medical Equipment (Trail)” and applying the American FLUKE ventilators gas flow detector (VT-PLUS), the data of ventilators, including the tidal volume of ventilators, the ratio of forced ventilation frequency and respiration, the concentration of inspired oxygen, the suction pressure, the PEEP levels, and other parameters, were measured.Results According to test and maintain all ventilators of our hospital from 2010 to 2014, it is clear to understand the working status of ventilators and maintain the quality problems of ventilators, as well as improving safety in use of ventilators and guaranteeing the requirement of clinical departments. Conclusion Detecting parameter values of ventilators have important clinical reference values to evaluate operating status of ventilators.关键词
呼吸机/潮气量/气体流量检测仪/质量控制Key words
ventilator/tidal volume/gas lfow detector/quality control分类
机械制造引用本文复制引用