中国药房2016,Vol.27Issue(1):140-142,3.DOI:10.6039/j.issn.1001-0408.2016.01.46
振荡器的使用对静脉用药调配中心百级净化台洁净度的影响
Effects of Oscillator on the Cleanliness of Class 100 Clean Bench in Pharmacy Intravenous Admixture Ser-vices
谢继青 1石继明 1赵文静 1于燕莉1
作者信息
- 1. 济南军区总医院药剂科,济南 250031
- 折叠
摘要
Abstract
OBJECTIVE:To study the influence of oscillator on the cleanliness of class 100 clean bench in PIVAS. METH-ODS:Using sedimentated bacteria and the number of dust particle as index,in common drug configuration room,antibiotics con-figuration room and risk drugs configuration room including biological safety cabinet and horizontal laminar flow,the cleanliness of class 100 clean bench were monitored when oscillator was set at clean bench and different positions in work and non-working state. RESULTS & CONCLUSIONS:In working and non-working state of oscillator,there was no difference in sedimentated bacteria and the number of dust particle which was in line with the requirements of 2010 edition of GMP,i.e. the application and location of oscillator didn't influence the cleanliness of class 100 clean bench. From a view of safety,it is suggested to place the oscillator in the left(or right)posterior wall of clean table when biological safety cabinets is used to dispense antibiotic and risk drugs.关键词
振荡器/净化台/洁净度/静脉用药调配中心Key words
Oscillator/Clean bench/Cleanliness/Pharmacy intravenous admixture services分类
医药卫生引用本文复制引用