原子能科学技术2016,Vol.50Issue(3):531-535,5.DOI:10.7538/yzk.2016.50.03.0531
组织等效正比计数器对单能中子响应的测量及模拟
Measurement and Simulation of TEPC’s Response to Mono-energetic Neutron
姜志刚 1王和义 1袁永刚 1陈华1
作者信息
- 1. 中国工程物理研究院 核物理与化学研究所,四川 绵阳 621900
- 折叠
摘要
Abstract
The tissue equivalent proportional counters are used as active neutron dose‐meters in radiation protection .A spherical TEPC was designed and employed for moni‐toring accelerator‐based mono‐energetic neutron sources . T he microdosimetric spectra of 0.36 ,0.65 ,0.8 ,1.0 and 1.3 MeV mono‐energetic neutrons were obtained .The absorbed dose ,mean quality factor and dose equivalent value were evaluated from the measured spectra .The comparison between H* (10) which was calculated based on the transforming factor of fluence to ambient dose equivalent and measured dose equivalent was conducted for each mono‐energetic neutron .The FLUKA simulation of TEPC was also implemented for comparison .It is found that the simulated data agree well with the experimental results and both reveal a good dose equivalent response of the spherical T EPC .关键词
TEPC/单能中子/微剂量谱/剂量当量响应/FLUKAKey words
T EPC/mono-energetic neutron/microdosimetric spectrum/dose equivalent response/FLUKA分类
能源科技引用本文复制引用
姜志刚,王和义,袁永刚,陈华..组织等效正比计数器对单能中子响应的测量及模拟[J].原子能科学技术,2016,50(3):531-535,5.