无线电工程2016,Vol.46Issue(5):95-98,4.DOI:10.3969/j.issn.1003-3106.2016.05.24
基于瞬时测频的频率捷变测试仪精度检定方法
Calibrating Method of Frequency Agility Instrument Accuracy Based on Instantaneous Frequency Measurement Technology
潘哲 1杜文超1
作者信息
- 1. 海军装备部驻天津军事代表局,北京100070
- 折叠
摘要
Abstract
Integrated tester of frequency agile radar is an important instrument for frequency agile seeker testing. Since available technology can’ t check the performance of dynamic frequency tracking,the paper adopts a system design method to solve the problem. The frequency agile testing requirements are analyzed,and a data⁃fitting IFM method is proposed,in which the precision and speed sat⁃isfy the requirement of tracking test.The system can check the performance of radar integrated tester effectively,and it has been imple⁃mented in engineering and achieved good effect.关键词
瞬时测频/导弹/频率捷变Key words
IFM/missile/frequency agile分类
信息技术与安全科学引用本文复制引用
潘哲,杜文超..基于瞬时测频的频率捷变测试仪精度检定方法[J].无线电工程,2016,46(5):95-98,4.