低MU对不同加速器X射线剂量学影响的研究

王刚 王惠 张宗春

中国医疗设备2016,Vol.31Issue(4):111-113,3.
中国医疗设备2016,Vol.31Issue(4):111-113,3.DOI:10.3969/j.issn.1674-1633.2016.04.029

低MU对不同加速器X射线剂量学影响的研究

Study on the Effect of X-ray Dosimetry of Different Accelerators at Low Monitor Units

王刚 1王惠 1张宗春1

作者信息

  • 1. 青岛市中心医院 放射物理科,山东青岛 266042
  • 折叠

摘要

Abstract

Objective By studying the effect of X-ray dosimetry of three accelerators, including Elekta Precise, Elekta Synergy, and Varian 21CX, this paper aimed to set up optimized parameters for treatment plan system.Methods On different level monitoring unit (MU), the research measured three basic parameters to describer X-ray characteristics: linearity of output dose, ifeld uniformity, and hardness. Results When 1≤MU<5, the dosage difference of each monitoring unit was less than 1.9%; when MU≥5, the dosage difference of each monitoring unit was less than 0.8%; when MU≥3, the errors of the uniformity, symmetry, and hardness were all within the allowable range.Conclusion Different types of occelerators had different parameters, the effects of low MU on accelerators varied. Thus it was necessary to measure basic characteristics of different accelerators at regular intervals.

关键词

低MU/输出剂量的线性/射野平坦度/射野对称性/射线质

Key words

low monitoring unit (MU)/linearity of output dose/ifeld uniformity/ifeld symmetry/actinogen