Objective:To evaluate the reliability of hematology analyzer on complete blood cell counting(CBC) and leukocyte differential count(DC).Methods: The performance of CBC and DC were evaluated by following guidelines and compared with that of Sysmex XE-2100 and manual operation.Results: The carryover rate, total repeatability and precision of BC-5000 could meet the standards, and the linearity was also reliable both in physiogenic and pathological conditions(r>0.998). Comparing with the results from XE-2100, the correlation was also satisfied(r>0.95). It has better repeatability in neutrophilic granulocytes, lymphocytes, eosinophile granulocyte and basophile granulocyte. The WBC samples with morphological performance were compared between BC-5000 and XE-2100. And the repeatability in neutrophilic granulocytes, lymphocytes and eosinophile granulocyte (r>0.96), the r value of monocyte was 0.9413 and basophil was 0.7270. While, the repeatability in BC-5000 and manual operation in neutrophilic granulocytes, lymphocytes and eosinophile granulocyte(r=0.9806,r=0.9691,r=0.9839), the r value of monocyte was 0.8993 and basophil was 0.4110.Conclusion: Except the suggestive diagnosis of abnormal leukocytes classification should be double checked by manual smear microscopic examination, the determination results from BC-5000 automatic hematology analyzer were accurate and reliable. And the main performance of BC-5000 could meet the clinical requirements.

关键词

血细胞分析仪/携带污染率/重复性/精密度/线性

Key words

Hematology analyzer/Carryover/Reproducibility/Precision/Linearity

分类

医药卫生

引用本文复制引用

刘晓婷,向代军,徐菡,乐家新,王成彬..迈瑞BC-5000全自动血细胞分析仪性能的可靠性研究[J].中国医学装备,2016,13(4):36-40,5.

基金项目

国家重大科学仪器设备开发专项(2012YQ18011708)“微生物快速检测技术在临床中的应用”;全军医学技术研究“十二五”重大专项研究课题(AWS11Z005-4)“师、团快速检验箱组” (2012YQ18011708)

中国医学装备

OACSTPCD

1672-8270

访问量0
|
下载量0
段落导航相关论文