中国医疗设备2016,Vol.31Issue(6):114-116,3.DOI:10.3969/j.issn.1674-1633.2016.06.031
Brilliance 16 CT机房的辐射防护设计与分析
Design and Analysis of Radiation Protection in Brilliance 16 CT Room
摘要
Abstract
Objective To design a Brilliance 16 CT room that meets the national standards in order to effectively protect the staff from the hazard of X ray radiation.Methods Firstly, the wall and ceiling shielding thickness of the room calculated according to the formula should be 4 mm and 3 mm lead equivalent respectively. Then, the wall, ceiling, protection door, observation window and ventilation system structure were respectively designed based on the computed results and the theoretical analysis.Results The Gamma radiation dose rate was monitored in ifve monitoring sites, including the observation window, control gate, room door, waiting area and control room. The maximum average annual effective dose for staff was about 0.0044 mSv, far below the state standard of 20 mSv. ConclusionThe research and design of the Brilliance 16 CT room can effectively protect the staff from the X ray radiation hazard.关键词
CT/辐射防护/屏蔽厚度/辐射剂量率Key words
CT/radiation protection/shielding thickness/radiation dose rate分类
医药卫生引用本文复制引用
曾嵘..Brilliance 16 CT机房的辐射防护设计与分析[J].中国医疗设备,2016,31(6):114-116,3.