中国医学装备2016,Vol.13Issue(6):30-32,33,4.DOI:10.3969/J.ISSN.1672-8270.2016.06.010
河北省医疗卫生机构生物安全柜性能抽样检测研究
The sampling test research on biological safety cabinet of medical and health institutions in Hebei province
摘要
Abstract
Objective:To understand the status of performance test of bio-safety cabinets in medical system in Hebei province.Methods:Stratified random sampling method was used in the performance test of 81 bio-safety cabinets in medical systems in Hebei province and statistics analysis was performed.Results: The qualified rate of noise was 98.77% and the illumination was 50.62%. The qualified rate of cleanliness and downdraft velocity were 53.09% and 60.49%. The qualified rate of integrality of HEPA filters was 79.01%.There was 0 bio-safety cabinet to be checked once a year. Conclusion: The qualified rate of integrality of filters, cleanliness and downdraft velocity were lower. Bio-safety cabinets were in annual check-up irregularly. So it was necessary to do a good job for management, maintenance and supervision of bio-safety cabinets.关键词
生物安全柜/防护性能/检测Key words
Bio-safety cabinets/Protecting performance/Test分类
医药卫生引用本文复制引用
师闻欣,卢安,崔泽,卢振敏,王茜,孙克勤,燕阔,张世联,张子翔,安鸿..河北省医疗卫生机构生物安全柜性能抽样检测研究[J].中国医学装备,2016,13(6):30-32,33,4.基金项目
河北省科技厅支撑计划(132777104D)“河北省二级以上医院生物安全柜检测与质控相关技术研究” (132777104D)