中国工程科学Issue(10):33-46,53,15.
FY-3B卫星微波湿度计热真空定标方法和结果分析
Thermal/vacuum calibration of microwave humidity sounder on FY-3B satellite
王振占 1张升伟 1李靖 1李芸 1吴琼1
作者信息
- 1. 中国科学院微波遥感技术重点实验室,中国科学院空间科学与应用研究中心,北京 100190
- 折叠
摘要
Abstract
The principle and methods on thermal/vacuum(T/V)testing of microwave humidity sounder (MWHS)on FY-3B satellite were described;correction algorithms on calibration errors were developed,and some main specifications of MWHS-3B,such as linearity,sensitivity and calibration precision,were obtained. The parameters derived from the T/V testing had been used in in-orbit data processing to produce brightness tem-perature products in FY-3B ground system.关键词
风云三号卫星/微波湿度计/发射前定标/热真空定标Key words
FY-3B satellite/microwave humidity sounder/prelaunch calibration/thermal/vacuum testing分类
信息技术与安全科学引用本文复制引用