基于ACR标准的MRI图像均匀度及层厚的检测
The detection of image uniformity and slice thickness of MRI system based on ACR ;standards
摘要
Abstract
Objective:To detect the image uniformity and slice thickness of Siemens Skyra 3.0 T and GE Excite HD 1.5 TMRI system based on ACR standards. Methods:Image uniformity and slice thickness of MRI were measured with ACR phantom, and the detection results were calculated. Results: The image uniformity were 96.0% and 92.7% of Siemens Skyra 3.0T and GE Excite HD 1.5T MRI respectively, the slice thickness were 4.64 mm and 5.58 mm, the thickness deviation is 0.36 mm and 0.58 mm, then the results measured in the research were in the normal range. Conclusion: The detection of image uniformity and slice thickness can effectively guarantee the image quality and the accuracy of slice thickness.关键词
磁共振成像系统/图像均匀度/层厚/ACR模体/质量控制Key words
MRI system/Image uniformity/Slice thickness/ACR phantom/Quality control分类
医药卫生引用本文复制引用
付丽媛,梁永刚,倪萍,陈自谦,陈坚,李威,林迪逵,刘冰川,陈建新..基于ACR标准的MRI图像均匀度及层厚的检测[J].中国医学装备,2016,13(7):16-19,4.基金项目
福建省科技计划(2016I0010)“基于同步视频脑电图、PET/CT和多模态功能磁共振对癫痫精准定位的应用研究”;军事医学计量科研专项课题(2011-JL2-014)“功能磁共振成像专用体模及质量控制检测方法的研究”;南京军区重大课题(14ZX23)“军人创伤后应激障碍的多模态功能磁共振研究” (2016I0010)