中国医学装备2016,Vol.13Issue(7):31-33,34,4.DOI:10.3969/J.ISSN.1672-8270.2016.07.010
新型微量注药泵与注射泵流速精度检测的对比与分析
A new detection and analysis of flow rate accuracy on a new tracetype micro-injection drug ;pump and syringe pump flow rate accuracy
摘要
Abstract
Objective:is toTo detect a new type of micro-injection pump, its flow rate of quality control testing and syringe pump through the comparative analysis, and study its accuracy and stability. Methods:A certain brand of injection flow rate accuracy and stability of drug pumps and syringe pumps were detected by the use of professional Fluke infusion device analyzer and a dedicated infusion line. Results:Compared the injection pump with syringe pump flow rate of errors, a flow rate was 1 ml/h, t. The errordeviation range of micro-injection pump flow rate is larger than the syringe pump, and a flow rate was 25 ml/h and 50 ml/h, both the same, . Wwith drug detection time went on, a flow rate of injection pump has tended to stabilize. After the flow rate is stable, the longer the detection time is, the overall error is smaller. At the time of 15 min, test data are in line with national testing standards. Conclusion:The combination of injection pump and infusion pump, the micro-injection pump can meet the growing demand for clinical departments, and its stability and accuracy can meet the national standards.关键词
注药泵/注射泵/流速精度/检测/质量控制Key words
Injection pump/Syringe pump/Velocity accuracy/Detect/Quality control分类
医药卫生引用本文复制引用
张华伟,柏正璐,蒋红兵..新型微量注药泵与注射泵流速精度检测的对比与分析[J].中国医学装备,2016,13(7):31-33,34,4.基金项目
南京市医学科技发展资金(QRX11033)“青年工程人才培养专项” (QRX11033)