摘要
Abstract
Objective To study the invitro interference of Breviscapine on common clinical biochemical test items .Methods Bre‐viscapine was added into normal pooled serum as experimental group ,while water for injection was added into normal pooled serum as control group .Hitachi 7600‐010 automatic biochemical analyzer was used to simultaneously test total bilirubin (TBIL) ,direct bili‐rubin(DBIL) ,total protein(TP) ,albumin(ALB) ,alanine aminotransferase(ALT) ,aspartate aminotransferase(AST) ,lactic dehy‐drogenase(LDH) ,urea(UREA) ,creatinine(CREA) ,uric acid(UA) ,total cholesterol(TCHO) ,triglyceride(TG) ,high density lipo‐protein cholesterol(HDL‐C) ,low‐density lipoprotein cholesterol(LDL‐C) ,α‐hydroxybutyrate dehydrogenase(α‐HBDH) ,creatine ki‐nase (CK) ,creatine kinase isozyme MB(CK‐MB) ,potassium(K) ,sodium(Na) ,chlorine(CL)and calcium(Ca) .Then the data were statistically analyzed .Results Breviscapine had the possibility of analytical interference on CREA ,UA ,TG ,TCHO ,HDL‐C and LDL‐C .Conclusion Breviscapine clould interfer the detection of CREA ,UA ,TG ,TCHO ,HDL‐C and LDL‐C ,which should be de‐tected before Breviscapine treatment or after the half‐life period of metabolism .关键词
灯盏花素/生化检验/干扰效应Key words
Breviscapine/biochemical test/interference effects