中国医疗设备2016,Vol.31Issue(7):26-29,4.DOI:10.3969/j.issn.1674-1633.2016.07.008
光学相干层析成像旋转失真评价研究
Study on the Evaluation of Rotational Distortion in Optical Coherence Tomography
摘要
Abstract
Optical coherence tomography (OCT) is an emerging branch of medical endoscopy, which can rapidly acquire high-resolution 3D images. The advancement of OCT catheter design enables OCT for the diagnostics of various diseases near tissue surface. The clinical translation of OCT is accelerating in many ifelds such as cardiovascular diseases, gastrointestinal diseases, pulmonary diseases and urinary diseases. Rotation distortion (RD) is a common problem with OCT endoscopy, which reveals the mechanical properties of the catheter and may severely deteriorate the integrity of images. Currently, no standard provides deifnition or testing method for RD. In this paper, the characteristic of RD was discussed and a quantitative method to evaluate RD was proposed. The results provided useful reference for the quality control of OCT endoscopy.关键词
医用内窥镜/光学相干层析成像/旋转失真/图像处理/质量控制Key words
medical endoscopy/optical coherence tomography/rotational distortion/image processing/quality control分类
机械制造引用本文复制引用
王浩,孟祥峰,刘艳珍,任海萍..光学相干层析成像旋转失真评价研究[J].中国医疗设备,2016,31(7):26-29,4.基金项目
国家科技支撑计划项目(2015BAI43H00),中国食品药品检定研究院中青年基金项目(2015C02)。 ()