中国医学装备2016,Vol.13Issue(8):31-34,4.DOI:10.3969/J.ISSN.1672-8270.2016.08.009
Sysmex XN20 A1全自动血细胞分析仪性能评价
Evaluation for the performance of Sysmex XN 20 A1 automation blood cell analyzer
金芳 1龙建成 2张春霞3
作者信息
- 1. 首都医科大学附属北京儿童医院检验中心 北京 100400
- 2. 滕州市中心人民医院器械科 山东 枣庄 277500
- 3. 军事医学科学院实验仪器厂 北京 100850
- 折叠
摘要
Abstract
Objective:To evaluate the performance for the Sysmex XN20 A1 automation blood cell analyzer.Methods: The precision, contaminative rate, linear range, blank, accuracy and various sample models were verified, and low values of blood platelet were compared with the methods of microscopy and dye.Results:The contaminative rate was lower than 0.4%. The linear arrange and precision were well for the analyzer. The values of RBC, HGB and PLT were within the range of 1±0.05 and the correlation coefficients were higher than 0.975. The bias of average and constant values by the accuracy verification samples test met the demands of accuracy. And the relative differences of various sample models meet the requirements of the comparability. The variable coefficient of low values of PLT was lower than 4% by the dye method.Conclusion: The Sysmex XN20 A1 automation blood cell analyzer has the characteristics of perfect precision, accuracy, low contaminative rate, broad linear arrange and good repeatability for the low level PLT. It can be applied in clinical laboratory.关键词
Sysmex XN全自动血细胞分析仪/性能评价/精密度/可比性/携带污染率Key words
Sysmex XN automation blood cell analyzer/Performance evaluation/Precision/Comparability/Contaminative rate分类
医药卫生引用本文复制引用
金芳,龙建成,张春霞..Sysmex XN20 A1全自动血细胞分析仪性能评价[J].中国医学装备,2016,13(8):31-34,4.