原子能科学技术2016,Vol.50Issue(9):1680-1687,8.DOI:10.7538/yzk.2016.50.09.1680
批制辐射加工用辐射变色薄膜剂量性能研究
Study on Radiochromic Film by Batch Production for Radiation Processing
陈义珍 1罗瑞 1陈克胜 1张卫东 1李明 1叶宏生 1林敏 1徐利军 1夏文 1崔莹1
作者信息
- 1. 中国原子能科学研究院计量与校准技术重点实验室,北京 102413
- 折叠
摘要
Abstract
The radiochromic film dosimeter was prepared using nylon as substrate and pararosaniline cyanide as dye in this paper .In 60 Co γ‐ray and 10 MeV electron beam ref‐erence radiation fields ,dosimetry response performance of the radiochromic film was studied .T he results show that the dosimeter has the response curve of good linearity in the dose range of 1‐210 kGy ,and no distinct dependence on energy and dose rate in the experiment .The repeatability of the radiochromic film dosimeter is better than 1.0% . The signal of radiochromic film dosimeter after irradiation is stable under the condition of low temperature storage within 2 weeks .In the dose range of 1‐210 kGy ,the exten‐sive uncertainty of the radiochromic film dosimeter is 4.4% (k=2) for 60 Co γ‐ray irradi‐ation and 6.2% (k=2) for electron beam irradiation .In addition ,this dosimetry system was used to measure depth‐dose distribution curve for 10 MeV electron beam ,which show s it’s suitable and practicable for electron beam irradiation .关键词
辐射加工/辐射变色薄膜/吸收剂量/γ射线/电子束Key words
radiation processing/radiochromic film/absorbed dose/γ-ray/electron beam分类
医药卫生引用本文复制引用
陈义珍,罗瑞,陈克胜,张卫东,李明,叶宏生,林敏,徐利军,夏文,崔莹..批制辐射加工用辐射变色薄膜剂量性能研究[J].原子能科学技术,2016,50(9):1680-1687,8.