首页|期刊导航|国际医药卫生导报|一台血凝分析仪维修后的性能验证

一台血凝分析仪维修后的性能验证

胡志清 侯小霞

国际医药卫生导报2016,Vol.22Issue(20):3161-3163,3.
国际医药卫生导报2016,Vol.22Issue(20):3161-3163,3.DOI:10.3760/cma.j.issn.1007-1245.2016.20.037

一台血凝分析仪维修后的性能验证

Performance evaluation of one coagulation analyzer after maintenance

胡志清 1侯小霞1

作者信息

  • 1. 471009 郑州大学附属洛阳中心医院检验科
  • 折叠

摘要

Abstract

Objective To evaluate the performance of one coagulation analyzer after maintenance.Methods The precision,linearity,and carry over were verified in turn according to EP5A2 and EP6-A2 of clinical and laboratory standards institute(CLSI).Results For the maintained coagulation analyzer,the normal and pathological values of relative intra-precision were 0.95% and 0.65% for prothrombin time (PT),0.57% and 0.62% for activated partial thromboplastin time (APTT),and 0.66% and 1.08% for fibrinogen(FIB),respectively.The normal and pathological values of relative total precision were 1.6% and 2.04% for prothrombin time (PT),2.12% and 2.06% for activated partial thromboplastin time(APTT),and 3.2% and 3.42 % for fibrinogen(FIB),respectively.The intra-precision and total precision values on the normal value of thrombin time (TT) were 0.96% and 2.73%.The linear regression equation of FIB was y=1.029x-0.1548,r2=0.9999.The carry over was 6.5%.Conclusions The precision,linearity,and carry over of the maintained coagulation analyzer meet the performance requirements of clinical laboratory.

关键词

性能验证/精密度/线性/携带污染率/血凝分析仪

Key words

Performance evaluation/Precision/Linear/Carry over/Coagulation analyzer