| 注册
首页|期刊导航|原子能科学技术|大面积气体闪烁正比计数器的研制及性能测试

大面积气体闪烁正比计数器的研制及性能测试

骆志平 庞洪超 刘阳 黄金峰 王莹 熊文俊

原子能科学技术2016,Vol.50Issue(10):1871-1874,4.
原子能科学技术2016,Vol.50Issue(10):1871-1874,4.DOI:10.7538/yzk.2016.50.10.1871

大面积气体闪烁正比计数器的研制及性能测试

Developement and Performance Testing of Large Surface Gas Scintillation Proportional Counter

骆志平 1庞洪超 1刘阳 1黄金峰 1王莹 1熊文俊1

作者信息

  • 1. 中国原子能科学研究院辐射安全研究所,北京 102413
  • 折叠

摘要

Abstract

In this work ,a gas scintillation proportional counter (GSPC) with effective detection area of 300 cm2 was built .The wavelength shifting material was used to couple the photomultiplier and scintillator to achieve the light collection . The performance testing was done to test the detector .The result shows that the energy resolution for 109Cd 22.6 keV X ray is 7.5% ,and the total detection efficiency is 12.4% when the source‐detector separation distance is 3.2 cm .The detector studied in this work could be used to measure the low energy X and γ radioactive sample in radiation contaminated facilities ,and to measure the internal intake activity of transuranium elements .

关键词

气体闪烁正比计数器/波长转换材料/能量分辨率/探测效率

Key words

gas scintillation proportional counter/wavelength shifting material/energy resolution/detection efficiency