广西医用诊断X射线摄影机质量控制检测结果及分析

梁桂强 覃志英 谢萍 梁挺 陈掌凡 陈发想 董颖

中国医学装备2016,Vol.13Issue(10):18-21,4.
中国医学装备2016,Vol.13Issue(10):18-21,4.DOI:10.3969/J.ISSN.1672-8270.2016.10.006

广西医用诊断X射线摄影机质量控制检测结果及分析

The results analysis of quality control detection on medical X-ray diagnostic equipment in Guangxi

梁桂强 1覃志英 1谢萍 1梁挺 1陈掌凡 1陈发想 1董颖1

作者信息

  • 1. 广西壮族自治区疾病预防控制中心放射卫生防护所 广西 南宁 530028
  • 折叠

摘要

Abstract

Objective:To detect image quality control in Guangxi medical X-ray diagnostic equipment in 2013, and analyze the factors of unqualified in quality control testing project, then providing a scientific basis to improve performance qualification rate of medical X-ray diagnostic equipment.Methods: The image quality control of 282 medical X-ray diagnostic equipment from 171 hospitals in 2013 were conducted through Barracuda x-ray output evaluation device, according to national specifications for testing of image quality control in medical X-ray diagnostic equipment (WS 76-2011).Results: In 282 medical X-ray diagnostic equipment, 235 are completely qualified and the qualified rate is 83.3%. The qualified rate of tertiary hospitals, second-class hospitals, and below second-class hospitals are 93.7%, 97.1%, 68.1%, respectively. In addition, the performance qualified rate of import machine is higher than the homebred machine. Conclusion: The performance and quality test qualified rate of x-ray imaging medical machine is low in below second-class hospitals. It needs to be improved on the maintenance, performance measuring and image quality control of medical X-ray diagnostic equipment.

关键词

X射线摄影机/质量控制/性能检测/结果分析

Key words

Medical X-ray diagnostic equipment/Quality control/Performance test/Result analysis

分类

医药卫生

引用本文复制引用

梁桂强,覃志英,谢萍,梁挺,陈掌凡,陈发想,董颖..广西医用诊断X射线摄影机质量控制检测结果及分析[J].中国医学装备,2016,13(10):18-21,4.

基金项目

广西自然科学基金(2014GXNSFAA118223)“广西医用辐射危害评价与控制技术研究”广西自然科学基金(2014GXNSFBA118173)“广西介入放射学工作人员和患者辐射防护状况及其控制策略研究” (2014GXNSFAA118223)

中国医学装备

OACSTPCD

1672-8270

访问量0
|
下载量0
段落导航相关论文