| 注册
首页|期刊导航|电气技术|无外接电源医疗设备静电问题处理方法分析

无外接电源医疗设备静电问题处理方法分析

张涛 杨雄 赵阳 周孟夏 夏清

电气技术Issue(11):106-110,5.
电气技术Issue(11):106-110,5.

无外接电源医疗设备静电问题处理方法分析

The Analysis of No External Power Supply Medical Equipment’s Electrostatic Problem

张涛 1杨雄 1赵阳 1周孟夏 1夏清1

作者信息

  • 1. 南京师范大学江苏省电气装备电磁兼容工程实验室,南京 210043
  • 折叠

摘要

Abstract

Since the electromagnetic compatibility standard YY0505-2012 was executed in the medical device industry, the electromagnetic compatibility of medical equipment has received much attention.Some common empirical measures about ESD protection have been taken, such as covering the equipment with a metal shell linked to the ground for shielding and using ESD devices on chip. However, there are rare studies involving no external power supply equipment. Therefore, this paper has explored and analyzed the problem of ESD in no external power supply medical equipment and has proposed three rectification measures for this problem with analyses of rectification cases. Rectification results indicate that the three ESD protective measures proposed in the paper are practical and have certain reference value for the research of ESD protection measures.

关键词

无外接电源/静电干扰模型/ESD防护措施/整改分析

Key words

no external power supply equipment/ESD interference model/ESD protective measures/rectification analysis

引用本文复制引用

张涛,杨雄,赵阳,周孟夏,夏清..无外接电源医疗设备静电问题处理方法分析[J].电气技术,2016,(11):106-110,5.

基金项目

江苏省自然科学基金(BK2011789)江苏省教育厅高校科研成果产业化推进项目 ()

电气技术

1673-3800

访问量0
|
下载量0