电气技术Issue(11):106-110,5.
无外接电源医疗设备静电问题处理方法分析
The Analysis of No External Power Supply Medical Equipment’s Electrostatic Problem
摘要
Abstract
Since the electromagnetic compatibility standard YY0505-2012 was executed in the medical device industry, the electromagnetic compatibility of medical equipment has received much attention.Some common empirical measures about ESD protection have been taken, such as covering the equipment with a metal shell linked to the ground for shielding and using ESD devices on chip. However, there are rare studies involving no external power supply equipment. Therefore, this paper has explored and analyzed the problem of ESD in no external power supply medical equipment and has proposed three rectification measures for this problem with analyses of rectification cases. Rectification results indicate that the three ESD protective measures proposed in the paper are practical and have certain reference value for the research of ESD protection measures.关键词
无外接电源/静电干扰模型/ESD防护措施/整改分析Key words
no external power supply equipment/ESD interference model/ESD protective measures/rectification analysis引用本文复制引用
张涛,杨雄,赵阳,周孟夏,夏清..无外接电源医疗设备静电问题处理方法分析[J].电气技术,2016,(11):106-110,5.基金项目
江苏省自然科学基金(BK2011789)江苏省教育厅高校科研成果产业化推进项目 ()