| 注册
首页|期刊导航|同位素|溢流式驻极体水氡浓度测量方法与装置

溢流式驻极体水氡浓度测量方法与装置

张辉 邓君 徐乐昌 李先杰

同位素2016,Vol.29Issue(4):241-247,7.
同位素2016,Vol.29Issue(4):241-247,7.DOI:10.7538/tws.2016.29.04.0241

溢流式驻极体水氡浓度测量方法与装置

The Approach and Device for Water Radon Measurement with Overflowing Electret

张辉 1邓君 2徐乐昌 1李先杰1

作者信息

  • 1. 核工业北京化工冶金研究院 北京 101149
  • 2. 中国疾病预防控制中心 辐射防护与核安全医学所 北京 100088
  • 折叠

摘要

Abstract

To solve the difficulties in “original”radon concentration measuring,this paper proposed an overflow type electret radon concentration measurement method and device.A 4 L jar was applied as ionization chamber and E-perm electret was applied as radon detector.When the steady radon contaminated water flowed through the ioniza-tion chamber,water radon was released with the nitrogen aerating disturbance.The average radon concentration was detected by the E-perm electret.Furthermore,RAD7-H2O kit was used to measure the radon concentration for comparison.The results showed that different methods had well consistency with the water radon concentration in the range of 4 to 97 Bq/L.

关键词

驻极体/水氡/辐射防护

Key words

electret/radon in water/radiation protection

分类

数理科学

引用本文复制引用

张辉,邓君,徐乐昌,李先杰..溢流式驻极体水氡浓度测量方法与装置[J].同位素,2016,29(4):241-247,7.

同位素

OACSTPCD

1000-7512

访问量3
|