原子能科学技术2016,Vol.50Issue(11):1949-1955,7.DOI:10.7538/yzk.2016.50.11.1949
基于制备型气相色谱的氙氡分离技术研究
Study on Decontaminating Radon from Xenon by Preparative Gas Chromatography
陈占营 1刘蜀疆 2王建龙 3常印忠 1樊元庆 4李奇3
作者信息
- 1. 清华大学核能与新能源技术研究院,先进核能技术协同创新中心,北京 100084
- 2. 禁核试北京国家数据中心和放射性核素实验室,北京 100085
- 3. 禁核试北京国家数据中心和放射性核素实验室,北京 100085
- 4. 清华大学放射性废物处理北京市重点实验室,北京 100084
- 折叠
摘要
Abstract
Radioxenon isotopes detection is of great interest for monitoring compliance with the comprehensive nuclear test ban treaty (CTBT ) ,and is of particular importance for the safe operations of nuclear facilities ,enviromental radiation survey and nuclear emergency monitoring .Since the daughters of radon including 214 Bi and 214 Pb ,can affect radioxenon measurement , radon should be removed before xenon measurement . T he study was carried out based on the preparative gas chromatography method .In this work ,an experimental setup for preparative gas chromatography was firstly estab‐lished .Utilizing the setup ,the xenon prepared performance and the radon decontami‐nate coefficient were tested separately .The results show that the experimental setup and the method for xenon preparation and radon decontaminating are practical and efficient .T he purified xenon volume ,the xenon concentration ,the xenon recovery and the radon decontaminate coefficient for the prepared sample are more than 11.5 mL (@ STP) ,80% ,95% and 105 ,respectively .关键词
制备型气相色谱/氙/氡/去污因子Key words
preparative gas chromatography/xenon/radon/decontaminate coefficient分类
化学化工引用本文复制引用
陈占营,刘蜀疆,王建龙,常印忠,樊元庆,李奇..基于制备型气相色谱的氙氡分离技术研究[J].原子能科学技术,2016,50(11):1949-1955,7.