Objective:To explore the technological method for calibrating and adjusting position deviations of X-ray pathway and mechanic fixed parts of X-ray machines, the X-ray light and X-ray irradiation field&field center position offset four sides in line according with the National Standard. Methods: We use a dedicated alignment module by repeating the correction, and ultimately achieve the desired objectives. Results:Ray will eventually be adjusted to the position of the four sides of the center&the standard position, and the radiation dose values are in line with GB. Conclusion:Equipment after calibration has been greatly improved in terms of quality image output, ray quality control in line with industry standards, and it improves the quality of clinical diagnosis of the diseases.

关键词

X射线机/照射野/位置偏离/校准

Key words

X-ray machine/Irradiation field/Position offset/Calibration

分类

医药卫生

引用本文复制引用

魏坤,余华良,谢杲..X射线机灯光照射野与X射线照射野中心点及四边偏离的调整校准[J].中国医学装备,2016,13(11):28-30,3.