中国医学装备2017,Vol.14Issue(1):15-17,3.DOI:10.3969/J.ISSN.1672-8270.2017.01.005
2016年西藏自治区日喀则市医用X射线诊断设备质量控制抽样检测调查与分析
Investigation and analysis on quality control sampling inspection of medical X-ray diagnostic equipment in Shigatse, Tibet Autonomous Region in 2016
摘要
Abstract
Objective:To investigate and analyze the status of quality control of medical X-ray diagnostic equipment in Shigatse, and propose recommendations for quality improvement. Methods: Applying quality control tests to all the medical X-ray diagnostic equipment in the sampled hospitals and analyzing the results by statistic method.Results: The qualification rates of equipment for fluoroscopy, radiography and computed tomography are 80%, 40% and 50%, respectively. The qualification rates is 57.1% in all of different levels of the hospitals. Conclusion: The status of quality control of medical X-ray diagnostic equipment in Shigatse is grasped. Both of the testing of quality control and the training of the technician should be strengthened in the future, all over Tibet Autonomous Region.关键词
质量控制/质量检测/医用X射线诊断设备/西藏自治区Key words
Quality control/Quality test/Medical X-ray diagnostic equipment/Tibet autonomous region分类
医药卫生引用本文复制引用
黄卓,徐辉,王建超,程金生,王爱萍,晋扎..2016年西藏自治区日喀则市医用X射线诊断设备质量控制抽样检测调查与分析[J].中国医学装备,2017,14(1):15-17,3.基金项目
国际原子能机构技术合作项目(CPR6006)“现代放射诊疗中医疗照射的质量保证” (CPR6006)