中国听力语言康复科学杂志2017,Vol.15Issue(1):17-20,4.DOI:10.3969/j.issn.1672-4933.2017.01.005
诺尔康人工耳蜗语前聋效果配对研究初步报告
Preliminary report of a matched case-control study of Nurotron cochlear implants
彭皎皎 1郑虹 1吴琦 1孔维丽1
作者信息
- 1. 四川大学华西医院耳鼻咽喉头颈外科 成都 610041
- 折叠
摘要
Abstract
Objective To evaluate and compare the rehabilitation of the prelingually profound deaf patients using domestic Nurotron cochlear implant and the imported cochlear implant.Methods 26 pairs of patients were retrospectively included who had the cochlear implantations during Jun 2012 to Nov 2015 with a six-months-follow-up at least. 26 patients with Nurotron cochlear implants were divided into the research group,while the 26 patients with the imported implants as the control group, whom were matched by the date of surgery, gender, age, medical history, preoperative language ability, et al. CAP and SIR were assessed to evaluate the effect of the implants.Results The average CAP and SIR score of the research group were 4.31±1.12 and 2.54±0.71, while of the control group were 4.85±1.19 and 2.88±1.07. There was no statistically significant difference (P >0.05) between neither the CAP nor the SIR scores of the two groups. 84.6%(22/26) patients of both group had CAP score ≥4, who were able to distinguish the speech. 57.7%(15/26) patients of the research group and 69.2%(18/26) of the control group with SIR score ≥3 had a degree of speech intelligibility.Conclusion There was no significant difference of the effects between the domestic Nurotron cochlear implants and the imported implants, and the domestic Nurotron cochlear implants can help the deafened patients effectively.关键词
诺尔康人工耳蜗/配对研究/语前聋/CAP/SIRKey words
Nurotron cochlear implant/Matched case-control study/Prelingually deaf/CAP/SIR引用本文复制引用
彭皎皎,郑虹,吴琦,孔维丽..诺尔康人工耳蜗语前聋效果配对研究初步报告[J].中国听力语言康复科学杂志,2017,15(1):17-20,4.