床边X线机摄影公众辐射剂量计算及辐射防护评价

储晓阳 柏彭开

中国医疗设备2017,Vol.32Issue(3):89-92,4.
中国医疗设备2017,Vol.32Issue(3):89-92,4.DOI:10.3969/j.issn.1674-1633.2017.03.024

床边X线机摄影公众辐射剂量计算及辐射防护评价

Public Radiation Dose Algorithm and Radiation Protection Assessment of Bedside X-ray Machine

储晓阳 1柏彭开1

作者信息

  • 1. 广西医科大学第二附属医院 医疗设备科,广西 南宁 530021
  • 折叠

摘要

Abstract

Objective To explore the radiation dose the public received while the mobile bedside X-ray machine is working. Methods Firstly, the radiation dose of hospital's centralized ward was detected. Based on the detected radiation dose rate, a mathematical model was built to transform it to the annual effective dose according to the clinical regularities. Then radiation protection assessment was made by comparing it with the radiation safety limit specified by the state. Results The annual radiation dose public received was far below than that of annual effective dose limit specified by the state when bedside X-ray machine was working, which met the national safety standards. Conclusion Though the public radiation dose of bedside X-ray machine is within the safety range, it is still necessary to implement protective measures in pediatric intensive care unit, neonatology department and patients' family members who stay beside patient.

关键词

移动式床边X线机/辐射防护/公众辐射剂量/剂量限值/剂量检测/剂量算法/防护评价

Key words

mobile bedside X-ray machine/radiation protection/public radiation dose/dose limit/dose detection/dose algorithm/calculation method of radiation/protection assessment