Abstract
Objective:The temperature and quality control monitoring methods of pressure steam sterilizer were analyzed to solve the sterilization temperature and sterilization time deviation of the sterilizer.Method: The external temperature and pressure detector to detect the performance of sterilization equipment: pressure steam sterilizer should use the temperature detector on the sterilization process temperature and time to detect the pressure steam sterilizer use temperature and pressure detector on the Bacteria program temperature, pressure and time to detect.Results: The temperature sensor drift caused by long-term use of the temperature sensor, the average deviation of 3.1%.Conclusion: The application of new external temperature and pressure detector on the performance of sterilization equipment can accurately monitor the temperature and other indicators. The temperature and pressure gauge inside the pressure steam sterilizer shall be monitored regularly or verified by the external temperature and pressure testing equipment to assess the accuracy of the temperature and pressure gauge of the pressure steam sterilization equipment.关键词
蒸汽灭菌器/质量控制/安全/温度/压力/检测Key words
Steam sterilizer/Quality control/Safety/Temperature/Pressure/Testing分类
医药卫生引用本文复制引用