摘要

Abstract

To study the influence of insulated container wall on measured phase of magnetic induction tomography(MIT).All of the containers were made of plexiglass which was insulated material.Saturated CaSO4 solution was injected into the containers.The containers were immerged into saturated CaSO4 solution in a cylindrical tank physical phantom.The phase shifts were measured before and after containers were immerged into.Two kinds of plexiglass containers were selected and two experiments were completed respectively.On the other hand, one solid plexiglass stick was immerged into saturated CaSO4 solution.The phase shifts were measured once per 1cm.The plot of phase shift with immerged depth was drawn.The ratio of the volume of two plexiglass container walls and that of the MIT phase shifts were closed.The plot of phase shift with immerged depth was approximately linear.The insulated plexiglass container wall is able to affect the measured phase data of MIT.Reducing the volume of container wall will mitigate the effect.

关键词

磁感应断层成像/绝缘材料/相位检测/物理模型/相关系数

Key words

Magnetic induction tomography/Insulated material/Phase detection/Physical phantom/Correlation coefficient

分类

医药卫生

引用本文复制引用

殷悦,刘锐岗..绝缘性容器壁对磁感应断层成像相位检测的影响研究[J].生物医学工程研究,2017,36(1):6-9,4.

基金项目

国家自然科学基金资助项目(61271101). (61271101)

生物医学工程研究

OACSTPCD

1672-6278

访问量0
|
下载量0
段落导航相关论文