| 注册
首页|期刊导航|物理学报|基于可调谐二极管激光技术利用小波去噪在2.008 μm波段对δ13CO2的研究

基于可调谐二极管激光技术利用小波去噪在2.008 μm波段对δ13CO2的研究

牛明生 王贵师

物理学报2017,Vol.66Issue(2):132-140,9.
物理学报2017,Vol.66Issue(2):132-140,9.DOI:10.7498/aps.66.024202

基于可调谐二极管激光技术利用小波去噪在2.008 μm波段对δ13CO2的研究

The research of δ13CO2 by use of wavelet de-noising at 2.008 μm based on tunable diode laser absorption spectroscopy

牛明生 1王贵师2

作者信息

  • 1. 曲阜师范大学激光研究所,山东省激光偏光与信息技术重点实验室,曲阜 273165
  • 2. 中国科学院安徽光学精密机械研究所大气物理化学研究二室,合肥 230031
  • 折叠

摘要

Abstract

Development of optical isotope techniques has provided scientists with a set of powerful tools for investigating the sources and sink of atmospheric CO2.Here we describe a continuous,high precision,compact and portable carbon dioxide isotope ratio laser multi-pass cell spectrometer with a tunable distribute feedback laser at 2.008 μm based on tunable diode laser absorption spectroscopy and,the spectrometer has good temperature and pressure stability.In order to deduce the noise,drift effect and background changes associated with low level signals,a superior signal processing technique of wavelet denoising,which possesses multi-level analytical resolutions both in time and frequency-domains,is introduced.After evaluating the method,evaluation ability and applicabilities of several common wavelet functions are analyzed and tested,the wavelet function of Haar is selected as an optimal wavelet basis function.Based on the analysis of the optimal decomposition level of Haa wavelet function,the VISU function is selected as an optimal wavelet threshold function.The denoising effect and measurement precision are evaluated by use of the VISU threshold function in the measurement process of carbon dioxide stable isotope ratio.The measurement results of carbon dioxide stable isotope ratio before and after suppressing the noises are compared in the same experiment conditions and,the inconsistent reasons of the measured results are theoretically analyzed.This technique allows the measurement of the δ-value for carbon dioxide isotopic ratios with a precision of-12.5‰ and the measuremnt results show that the wavelet denoising measuring results have higher measurement accuracy,and the measurement precise of carbon dioxide isotope ratio is 7.3 times the original measurement results.The application of the wavelet denoising to the carbon dioxide isotope ratio measurement for the first time proves that the capability of the new near-infrared direct absorption technique to measure isotope ratio can permit high-frequency,near-continuous isotope measurement and obtain the high precision and accurate real-time stable isotope data directly in the field.This technique provides an important tool for studying the resource and sink of green house gases in the future.

关键词

同位素测量/新型多通池/小波去噪/测量精度

Key words

isotope measurement/novel multi-pass cell/wavelet denoising/measurement precise

引用本文复制引用

牛明生,王贵师..基于可调谐二极管激光技术利用小波去噪在2.008 μm波段对δ13CO2的研究[J].物理学报,2017,66(2):132-140,9.

基金项目

国家自然科学基金(批准号:41405022)和曲阜师范大学博士启动基金(批准号:20130760)资助的课题.Project supported by the National Natural Science Foundation of China (Grant No.41405022) and the Qufu Normal University Fundation,China (Grant No.20130760). (批准号:41405022)

物理学报

OA北大核心CSCDCSTPCDSCI

1000-3290

访问量0
|
下载量0
段落导航相关论文