2种不同固定头架在头颈部肿瘤放疗中的摆位误差研究

胡燕 王亮和 陈星萌 邹雅 付敬国

医疗卫生装备2017,Vol.38Issue(3):80-82,93,4.
医疗卫生装备2017,Vol.38Issue(3):80-82,93,4.DOI:10.7687/J.ISSN1003-8868.2017.03.080

2种不同固定头架在头颈部肿瘤放疗中的摆位误差研究

Influences of two head holders on setup errors during head and neck neoplasms radiotherapy

胡燕 1王亮和 1陈星萌 1邹雅 1付敬国1

作者信息

  • 1. 430060武汉,武汉大学人民医院放疗科
  • 折叠

摘要

Abstract

Objective To compare the effects of two head holders on the setup errors during head and neck neoplasms radiotherapy.Methods Totally 30 patients were divided into groups A and B randomly and equally.Group A applied TypeSTM Extension for setup,and group B used Type-S Overlay Board,with thermoplastic film applied to fixation.Varian23IX image guided linear accelerator was involved in the therapy,and the patients underwent CBCT-guided therapy once a week in 6 weeks.The setup errors at left-right,head-foot and anterior-posterior directions were obtained by bone registration and manual checking.Results The absolute setup error at anterior-posterior direction of group A,was (2.31 ± 1.35) mm,which was significantly higher than (1.88±1.79) mm of group B (P=0.032).There were no obvious differences between the setup errors of group A and those of group B (P>0.05),and the weekly absolute errors of group A were not statistically different from those of group B (P>0.05).Conclusion There are no significant differences between the effects of the two head holders on the setup errors during head and neck neoplasms radiotherapy,and proper head holder can be selected according to clinical requirements.

关键词

锥形束CT/头颈部肿瘤/放射疗法/头颈肩固定板/摆位误差

Key words

CBCT/head and neck neoplasm/radiotherapy/head and neck positioning board/setup error

分类

医药卫生

引用本文复制引用

胡燕,王亮和,陈星萌,邹雅,付敬国..2种不同固定头架在头颈部肿瘤放疗中的摆位误差研究[J].医疗卫生装备,2017,38(3):80-82,93,4.

医疗卫生装备

OACSTPCD

1003-8868

访问量0
|
下载量0
段落导航相关论文