现代检验医学杂志2017,Vol.32Issue(2):160-164,5.DOI:10.3969/j.issn.1671-7414.2017.02.045
人来源样品中铝、砷、铬、镉和汞微量元素检测分析前因素的质量控制
Quality Control of Preanalytical Variations for the Determination of Trace Element Aluminum, Arsenic,Chromium, Cadmium and Mercury in Samples of Human Origin
摘要
Abstract
The aims of this article is to provide the quality control requirements of preanalytical variation for the determination of trace element aluminum,arsenic,chromium,cadmium and mercury in samples of human origin,reduce the influence of preanalytical variation on the test results.Refer to the Clinical and Laboratory Standards Institute (CLSI) documents,"Control of Preanalytical Variation in Trace Element Determinations" and other references and guidelines,the methods of quality control of aluminum,arsenic,chromium,cadmium and mercury determination have been made,including:the factors needed to be considered in collection,preservation,transportation and other preanalytical factors,the abilities and considerations of laboratory staff,etc.Hope this article provide some useful suggestions and help to the laboratories of determination of aluminum,arsenic,chromium,cadmium and mercury in samples of human origin.关键词
铝/砷/铬/镉/汞/人来源样品/分析前因素/污染控制Key words
aluminum/arsenic/chromium/cadmium/mercury/sample of human origin/preanalytical variation/contamination control分类
医药卫生引用本文复制引用
钟堃,王薇,何法霖,袁帅,王治国..人来源样品中铝、砷、铬、镉和汞微量元素检测分析前因素的质量控制[J].现代检验医学杂志,2017,32(2):160-164,5.基金项目
北京市自然科学基金资助项目(7143182). (7143182)