The detection of volatile phenol in sea water was optimized by continuous flow analyzer method.The results of this study could provide scientific basis for the determination of volatile phenol in sea water for scientific and technical personnel of testing agencies,and provide a scientific basis for the standardization of this method.The experimental results showed that this method had high precision and accuracy in the linear range of 0~0.100 mg/L,the minimum detection limit was 0.002 mg/L.The range of relative standard deviation was 2.21%~4.65%,which were all less than 5%(standard requirement).There was no significant difference between continuous flow analysis method and classical method.The method was fast,simple and environmental protection,and it would be worth to further study and application.关键词