| 注册
首页|期刊导航|中国医疗设备|双源CT双能量线性融合重建去除腰椎固定物金属伪影的初步研究

双源CT双能量线性融合重建去除腰椎固定物金属伪影的初步研究

李笑石 石磊 刘豹 常英娟 石明国 余厚军

中国医疗设备2017,Vol.32Issue(5):19-21,31,4.
中国医疗设备2017,Vol.32Issue(5):19-21,31,4.DOI:10.3969/j.issn.1674-1633.2017.05.005

双源CT双能量线性融合重建去除腰椎固定物金属伪影的初步研究

Preliminary Study of Utilizing Dual Source CT Dual Energy Linear Fusion Technology to Eliminate Lumbar Metal Implants Artifacts

李笑石 1石磊 1刘豹 1常英娟 1石明国 1余厚军1

作者信息

  • 1. 第四军医大学西京医院 放射科,陕西西安 710032
  • 折叠

摘要

Abstract

Objective To explore the clinical application value of the dual source CT dual energy scanning linear fusion technology for eliminating lumbar metal implants artifacts. Methods 33 patients that were going to check after lumbar metal implants were selected for our research. The dual source CT dual energy scanning with the voltages were separately 80 kV and 140 kV was used to gain the imaging data for linear fusion, which we marked as group A. The regular 120 kV imaging data was simulated, which we marked as group B. Multiple planar reconstruction, volume reality, maximum intensity projection, reconstruction image quality and the artifacts were evaluated. Results The images with high quality under linear fusion were 90% and 33%, respectively. The images with no artifact in group A and B accounted for 91% and 36%, respectively. The difference between group A (Z=5.74, P=0.00) and group B (Z=6.74, P=0.00) was statistically significant. Conclusion It was very accurate and effective for the dual source CT dual energy scanning linear fusion technology to eliminate lumbar metal implants artifacts. It could clearly show the location shape, and other microscopic structures of the lumbar metal fixture. In addition, it could clearly show right and left psoas muscles without artifacts interference. Therefore, the dual source CT dual energy scanning linear fusion technology can obtain high quality images and meet the diagnostic requirements.

关键词

腰椎金属固定物/计算机断层摄影术/X射线/放射摄影术/双能扫描投影/线性融合

Key words

lumbar metal fixture/computed tomography/X-ray/radiography/dual energy scanned projection/linear fusion

分类

医药卫生

引用本文复制引用

李笑石,石磊,刘豹,常英娟,石明国,余厚军..双源CT双能量线性融合重建去除腰椎固定物金属伪影的初步研究[J].中国医疗设备,2017,32(5):19-21,31,4.