| 注册
首页|期刊导航|中国医疗设备|国产动态心电记录仪CT-08的临床应用评价

国产动态心电记录仪CT-08的临床应用评价

毛彬 孙静 郑骏 陈华 王吉鸣 冯靖祎

中国医疗设备2017,Vol.32Issue(5):98-101,4.
中国医疗设备2017,Vol.32Issue(5):98-101,4.DOI:10.3969/j.issn.1674-1633.2017.05.025

国产动态心电记录仪CT-08的临床应用评价

Evaluation on ClinicalApplication of Dynamic ECG Recording Box CT-08

毛彬 1孙静 1郑骏 1陈华 1王吉鸣 1冯靖祎1

作者信息

  • 1. 浙江大学医学院附属第一医院 医学工程部,浙江 杭州 310006
  • 折叠

摘要

Abstract

Objective This study aimed to compare the performance of dynamic ECG recording box (Baihui CT-08) and conventional ECG machine in premature beat acquisition. Methods Premature beat signal from a physiological signal simulator was simultaneously collected via the 2 devices. A total of 72 sets of signal were obtained and statistically analyzed to verify the consistency of the 2 devices. Results No statistically significant difference was found between the 2 devices in the function of total beats and premature beats recording on all types of heart disorders except for the type of "PVC1 R on T, ventriculus sinister". Conclusion CT-08 is highly consistent with Holter in premature beat recording, making it possible to be promoted as a demonstration product in the project of national innovative medical devices and technical achievements transformation.

关键词

心电记录仪/早搏/心电图/心血管疾病

Key words

electrocardiograph recorder/premature beat/electrocardiography/cardiovascular disease

分类

机械制造

引用本文复制引用

毛彬,孙静,郑骏,陈华,王吉鸣,冯靖祎..国产动态心电记录仪CT-08的临床应用评价[J].中国医疗设备,2017,32(5):98-101,4.

基金项目

浙江省科技厅"国家创新医疗器械产品与技术成果转化工程"子课题(2013T301-16). (2013T301-16)

中国医疗设备

OACSTPCD

1674-1633

访问量0
|