中国医疗设备2017,Vol.32Issue(5):98-101,4.DOI:10.3969/j.issn.1674-1633.2017.05.025
国产动态心电记录仪CT-08的临床应用评价
Evaluation on ClinicalApplication of Dynamic ECG Recording Box CT-08
摘要
Abstract
Objective This study aimed to compare the performance of dynamic ECG recording box (Baihui CT-08) and conventional ECG machine in premature beat acquisition. Methods Premature beat signal from a physiological signal simulator was simultaneously collected via the 2 devices. A total of 72 sets of signal were obtained and statistically analyzed to verify the consistency of the 2 devices. Results No statistically significant difference was found between the 2 devices in the function of total beats and premature beats recording on all types of heart disorders except for the type of "PVC1 R on T, ventriculus sinister". Conclusion CT-08 is highly consistent with Holter in premature beat recording, making it possible to be promoted as a demonstration product in the project of national innovative medical devices and technical achievements transformation.关键词
心电记录仪/早搏/心电图/心血管疾病Key words
electrocardiograph recorder/premature beat/electrocardiography/cardiovascular disease分类
机械制造引用本文复制引用
毛彬,孙静,郑骏,陈华,王吉鸣,冯靖祎..国产动态心电记录仪CT-08的临床应用评价[J].中国医疗设备,2017,32(5):98-101,4.基金项目
浙江省科技厅"国家创新医疗器械产品与技术成果转化工程"子课题(2013T301-16). (2013T301-16)