医疗卫生装备2017,Vol.38Issue(4):102-104,3.DOI:10.7687/j.issn1003-8868.2017.04.102
GM260型便携式血糖仪与AU5821全自动生化分析仪比对分析
Comparative analysis of GM260 portable blood glucose meter and AU5821 automatic biochemical analyzer
谢强 1肖传宇 1陶峰 1王宁 1邱枫 1朱中梅 1王小艳 1谢守凤1
作者信息
- 1. 441200湖北枣阳,枣阳市第一人民医院检验科
- 折叠
摘要
Abstract
Objective To compare GM260 portable blood glucose meter and AU5821 automatic biochemical analyzer in order to prove the accuracy of GM260 and its applicability for clinical use.Methods Totally 20 pieces of EDTA-K2 anticoagulative specimens and 23 GM260 meters were numbered,and each specimen underwent examinations by both GM260 and AU5821,then the bias between the two kinds of devices was calculated.Results The maximal bias between GM260 and AU5821 was 0.47 mmol/L and all the meters had the bias between-0.83 and 0.83 mmol/L in case of 5 specimens with the glucose concentration less than 4.2 mmol/L;the maximal bias between GM260 and AU5821 was 18.07% and all the meters had the bias between-20% and 20% in case of 15 specimens with the glucose concentration not less than 4.2 mmol/L;the examination results by GM260 all accorded with industrial standard.The results by GM260 were lower than those by AU5821,and the maximal negative deviation was-13.43%.Conclusion Portable blood glucose meter can only be used for screening,and automatic biochemical analyzer is the preferred device for diabetes diagnosis.关键词
血糖仪/全自动生化分析仪/血糖/比对分析/糖尿病Key words
blood glucose meter/automatic biochemical analyzer/glucose/comparative analysis/diabetes分类
医药卫生引用本文复制引用
谢强,肖传宇,陶峰,王宁,邱枫,朱中梅,王小艳,谢守凤..GM260型便携式血糖仪与AU5821全自动生化分析仪比对分析[J].医疗卫生装备,2017,38(4):102-104,3.