医疗卫生装备2017,Vol.38Issue(4):105-107,3.DOI:10.7687/j.issn1003-8868.2017.04.105
尿液分析仪质控量值范围研究
Study on value range of urine analyzer quality control
雷勋祖 1牛国喻 1刘鹏 1胡松林2
作者信息
- 1. 541002广西桂林,解放军181医院医学工程科
- 2. 541002广西桂林,解放军181医院医学检验科
- 折叠
摘要
Abstract
Objective To study the reference range of the urine analyzer and the target value of the common quality control object.Methods The reference ranges of urine analyzer test items from different mnanufacturers were investigated,and laboratory bias proportion was used as an indicator of the evaluation of the target value scheme.EQA raw data of Guangxi clinical examination center in 2014 were analyzed.Results The target value ranges of pH,specific gravity,bilirubin and nitrite were found reliable,while the ones for WBC,URO,RBC,ketone body,glucose and protein needed improving.Conclusion The target values and evaluation indexes of EQA have to be reselected for mutual recognition of the results.关键词
尿液分析仪/质控/靶值/量值范围/临床检验Key words
urine analyzer/quality control/target value/value range/clinical laboratory examination分类
医药卫生引用本文复制引用
雷勋祖,牛国喻,刘鹏,胡松林..尿液分析仪质控量值范围研究[J].医疗卫生装备,2017,38(4):105-107,3.