根管冲洗液对RAYPEX 6及PROPEX Ⅱ电子根尖定位仪测量准确性的影响

胡玉凤 朱静 王丽琴 金文忠

临床口腔医学杂志2017,Vol.33Issue(5):289-291,3.
临床口腔医学杂志2017,Vol.33Issue(5):289-291,3.DOI:10.3969/j.issn.1003-1634.2017.05.011

根管冲洗液对RAYPEX 6及PROPEX Ⅱ电子根尖定位仪测量准确性的影响

The influence of root canal irrigation on the accuracy of RAYPEX 6 and PROPEX Ⅱ electronic apex locator

胡玉凤 1朱静 1王丽琴 1金文忠2

作者信息

  • 1. 上海市口腔医院,复旦大学附属口腔医院 上海 200031
  • 2. 上海交通大学医学院附属第九人民医院上海200011
  • 折叠

摘要

Abstract

Objective:To investigate the influence of three kinds of root canal irrigations on the accuracy of RAYPEX 6 and PROPEX Ⅱ electronic apex locator.Methods:RAYPEX 6 and PROPEX Ⅱ were selected as electronic apex locators.The root canal length of 40 extracted teeth with single straight root canal was measured in the case of 5.25%NaClO,0.9%NaCl and 3% H2O2.The difference between the electronic length and actual length was calculated.Statistic software SPSS 22.0 was used to assess the accuracy by the method of independent sample t test.Results:All mean difference value of these two electronic apex locators were less than 0.5 mm.There was no statistically significant difference between the two locators by using 0.9%NaCl(P>0.05).There was statistically significant difference in the case of 5.25%NaClO and 3% H2O2,respetively(P<0.05).Conclusion:Different types of root canal irrigation influenced the accuracy of RAYPEX 6 and PROPEX Ⅱ electronic apex locator,but the accuracy was still able to meet the requirements of clinical root canal length measurement.

关键词

电子根尖定位仪/根管冲洗液/根管长度

Key words

Electronic apex locator/Root canal irrigation/Root canal length

分类

医药卫生

引用本文复制引用

胡玉凤,朱静,王丽琴,金文忠..根管冲洗液对RAYPEX 6及PROPEX Ⅱ电子根尖定位仪测量准确性的影响[J].临床口腔医学杂志,2017,33(5):289-291,3.

基金项目

上海市卫生和计划生育委员会青年科研课题资助项目(20134Y029) (20134Y029)

临床口腔医学杂志

OACSTPCD

1003-1634

访问量0
|
下载量0
段落导航相关论文