计量学报2016,Vol.37Issue(z1):71-74,4.DOI:10.3969/j.issn.1000-1158.2016.z1.16
热释光剂量计刻度过程中有关电子平衡的Monte-Carlo模拟计算与实验验证
Monte-Carlo Simulation and Experimental about Electron Equilibrium for Calibration of Thermoluminescent Dosimeter
郝艳梅 1吴笛 1李德红 2黄建微 1党永乐 1李伦1
作者信息
- 1. 中国计量科学研究院,北京100029
- 2. 南华大学,湖南衡阳421001
- 折叠
摘要
Abstract
Using Monte-Carlo calculation and experimental measurement,the thickness of electron equilibrium compensating material of LiF (Mg,Cu,P) thermoluminescent dosimeter (TLD) under different X ray is studied.The stimulation result and the experimental result match each other very well,it shows that when the X ray effective energy is 83 keV,118 keV,208 keV,a certain thickness of electron equilibrium compensating material is needed on the surface of TLD,and the thickness increases with the X ray energy.关键词
计量学/热释光剂量计/电子平衡/蒙特卡洛模拟Key words
metrology/thermoluminescent dosimeter/electron equilibrium/Monte-Carlo simulation分类
通用工业技术引用本文复制引用