首页|期刊导航|中国医学装备|不同重建算法对低剂量头颈部CTA扫描图像质量的影响

不同重建算法对低剂量头颈部CTA扫描图像质量的影响

齐天白 侯珺琳

中国医学装备2017,Vol.14Issue(6):14-17,4.
中国医学装备2017,Vol.14Issue(6):14-17,4.DOI:10.3969/J.ISSN.1672-8270.2017.06.005

不同重建算法对低剂量头颈部CTA扫描图像质量的影响

Effects of different reconstruction algorithms on the image quality of CTA with low dose on head and neck

齐天白 1侯珺琳1

作者信息

  • 1. 湖州市第三人民医院设备科 浙江 湖州 313000
  • 折叠

摘要

Abstract

Objective:To discuss the effect of different reconstruction algorithms for the image quality of the computed tomographic angiography (CTA) with low dose on head and neck through comparing the images obtained from different reconstruction algorithms, which included filtering and reflection projection technique (FBP), adaptive statistical iterative (ASIR) model and based on iterative reconstruction (MBIR) after CTA with low dose was applied on head and neck.Methods:30 patients who were implemented the CTA on head and neck were enrolled in this study. The gem energy spectrum CT was applied to scan their head and neck, and the tube voltage was set in low dose (80 kVp), and 60-80 mL contrast agent (350 mg I/mL) was injected in elbow vein as 4mL/s. The other parameters were chosen as routine requirement. The obtained original images were reconstructed by using FBP, 60% ASIR and MBIR, and then the three kinds images were compared and scored by using relative statistical analysis.Results: For the using of different reconstruction algorithms (FBP, 60% ASIR and MBIR), the results of image noise showed MBIR<60% ASIR<FBP, and their differences were significant. The results of contrast noise ratio showed MBIR>60% ASIR>FBP, and the differences among them were significant. Finally, about the subjective score of image, the results showed MBIR>60% ASIR>FBP, and their differences also were significant.Conclusion: Under gem energy spectrum CT, the image noise of MBIR is lowest and its image quality is best, and it has diagnosis value and meaning in clinical practice.

关键词

宝石能谱CT/低剂量/头颈部血管成像/图像质量/重建方式

Key words

Energy spectrum CT/Low dose/Head and neck vascular imaging/Quality of image/Reconstruction method

分类

医药卫生

引用本文复制引用

齐天白,侯珺琳..不同重建算法对低剂量头颈部CTA扫描图像质量的影响[J].中国医学装备,2017,14(6):14-17,4.

中国医学装备

OACSTPCD

1672-8270

访问量0
|
下载量0
段落导航相关论文