电子器件2017,Vol.40Issue(3):737-741,5.DOI:10.3969/j.issn.1005-9490.2017.03.044
基于时差法和TDC-GP2的超声波热量表设计
Design of Ultrasonic Heat Meter Based on TimeDifference Method and TDC-GP2
崔永俊 1张彬彬 2杨兵1
作者信息
- 1. 中北大学电子测试技术国家重点实验室,太原 030051
- 2. 中北大学仪器科学与动态测试教育部重点实验室,太原 030051
- 折叠
摘要
Abstract
In order to achieve high precision of calorific value measurement,the design of ultrasonic heat measurement system is based on TDC-GP2.The system uses platinum resistance temperature sensor PT1000 and ultrasonic transducer as sensitive units,TDC-GP2 as testing unit,and the combination of FPGA and PC to realize the acquisition,processing and transmission of data.Through the test,the maximum measurement error is less than 3%,the maximum error of temperature measurement is less than 0.5 ℃,Calorific value measurement error is better than 0.3%.关键词
时差法/铂电阻温度传感器/超声波换能器/TDC-GP2/上位机Key words
time difference method/platinum resistance temperature sensor/ultrasonic transducer/TDC-GP2/pc分类
信息技术与安全科学引用本文复制引用
崔永俊,张彬彬,杨兵..基于时差法和TDC-GP2的超声波热量表设计[J].电子器件,2017,40(3):737-741,5.