| 注册
首页|期刊导航|中国医学装备|PET-CT受检者周围剂量水平检测与分析

PET-CT受检者周围剂量水平检测与分析

练德幸 张庆召 张奇 梁婧 朱卫国 葛丽娟 侯长松 张震

中国医学装备2017,Vol.14Issue(7):44-47,4.
中国医学装备2017,Vol.14Issue(7):44-47,4.DOI:10.3969/J.ISSN.1672-8270.2017.07.011

PET-CT受检者周围剂量水平检测与分析

The detection and analysis of dose level for ambient of the patient received PET-CT examination

练德幸 1张庆召 1张奇 1梁婧 1朱卫国 1葛丽娟 1侯长松 1张震1

作者信息

  • 1. 辐射防护与核应急中国CDC重点实验室 中国CDC辐射防护与核安全医学所 北京 100088
  • 折叠

摘要

Abstract

Objective: To grasp the level of dose equivalent rate for ambient of patients at different time and different distance after they were injected radiopharmaceuticals during the detection of PET-CT, and evaluate the key point of protection during the process of detection. Methods: 20 patients received detection of PET-CT were selected and a fast measurements platform was established. In the measurement, the real time data of portable instrument of γ ambient dose equivalent rate were sent to client of laptop by extended wire. And the results of detection could be real time recorded on remote. The ambient dose equivalent rates of different time, different direction and different distance were detected after the patients were injected 18F-FDG. Results: The range of ambient dose equivalent rate was 15.1μSv/h~33.9μSv/h when the distance was 1m between detected point and patients who were injected 18F-FDG, and then the rate decreased to 7.47μSv/h~20.2μSv/h after PET-CT detection finished. This meant its descend amplitude was 40.8%~78.3%. On the direction of detection, the ambient dose equivalent rate of prethorax and backside of patients were higher than the left and right side of them. For different distance, the ambient dose equivalent rate of 0.5m away from patients was rough 3 times of that of 1m away from patients. Conclusion: During the detection of PET-CT, the ambient dose equivalent was higher than normal level after patients were injected 18F-FDG, therefore, the protection for caretakers and the management for patients injected 18F-FDG should be strengthened so as to reduce the radiation dose on related personnel.

关键词

正电子发射计算机断层显像-CT/周围剂量当量率/辐射防护/18F-氟代脱氧葡萄糖

Key words

Positron emission tomography–computed tomography(PET-CT)/Ambient dose equivalent rate/Radiation protection/18F-flurodeoxyglucose

分类

医药卫生

引用本文复制引用

练德幸,张庆召,张奇,梁婧,朱卫国,葛丽娟,侯长松,张震..PET-CT受检者周围剂量水平检测与分析[J].中国医学装备,2017,14(7):44-47,4.

基金项目

国家重大科学仪器设备开发专项(2012YQ180118)"大批量人群核辐射体内污染快速检测仪" (2012YQ180118)

中国医学装备

OACSTPCD

1672-8270

访问量0
|
下载量0
段落导航相关论文