CT高级建模迭代重建算法对图像质量的影响初步研究

陆亚光 庄静文 吴天棋 白玫

中国医学装备2017,Vol.14Issue(7):68-71,4.
中国医学装备2017,Vol.14Issue(7):68-71,4.DOI:10.3969/J.ISSN.1672-8270.2017.07.017

CT高级建模迭代重建算法对图像质量的影响初步研究

Preliminary study on the effect of advanced modeled iterative reconstruction algorithm for image quality

陆亚光 1庄静文 1吴天棋 1白玫1

作者信息

  • 1. 首都医科大学宣武医院医学工程处 北京 100053
  • 折叠

摘要

Abstract

Objective: To research the effect of advanced modeled iterative reconstruction (ADMIRE) algorithm for images quality of multi-slice spiral CT scanner of SIEMENS. Methods: Under the same scanning condition, the CTP515 and CTP528 of Catphan500 were scanned by using spiral method of SOMATOM Force CT (SIEMENS). The images were reconstructed by two conditions (didn't use ADMIRE algorithm, or used different intensity of ADMIRE algorithm). And then, these images from different conditions were evaluated by subjective method and objective method, respectively. Results: When the low contrasts were 0.3%, 0.5% or 1.0%, the image noises were 4.90, 4.50 and 5.23, respectively, under didn't use ADMIRE algorithm. In the same contrast, the noises were reduced 18,35%, 53.23% and 18.82%, respectively, under ADMIRE2. And the noises were reduced 52.32%, 20.33% and 53.79%, respectively, under ADMIRE5. When ADMIRE0, ADMIRE2 and ADMIRE5 were used to reconstruct the image of CTP528 module, respectively, the smallest details that could be observed were identical that was 7.0 Line Pairs/cm under subjective observation for images. Conclusion: Using ADMIRE to reconstruct images can effectively reduce the noise while the image resolution don't be affected. And to enhance the resolution power of low contrast for CT image can more accurately help clinician to observe lesion with similar tissue density and contribute to diagnose disease.

关键词

多排螺旋CT/低对比度分辨率/高级建模迭代重建算法/主观评价/图像质量

Key words

Multislice spiral CT (MSCT)/Low contrast resolution/Advanced modeled iterative reconstruction (ADMIRE)/Subjective evaluation/Image quality

分类

医药卫生

引用本文复制引用

陆亚光,庄静文,吴天棋,白玫..CT高级建模迭代重建算法对图像质量的影响初步研究[J].中国医学装备,2017,14(7):68-71,4.

基金项目

国家自然科学基金(81372923)"宽探测器螺旋CT所致辐射剂量评价方法研究" (81372923)

中国医学装备

OACSTPCD

1672-8270

访问量0
|
下载量0
段落导航相关论文