常松涛 田棋杰 何锋赟 余毅 李周

物理学报2017,Vol.66Issue(15):52-63,12.
物理学报2017,Vol.66Issue(15):52-63,12.DOI:10.7498/aps.66.150701

基于球面反射温阑的红外探测器变f数设计

Design of varying f/number of cooled infrared detectors based on spherical reflecting warm shield

常松涛 1田棋杰 1何锋赟 2余毅 1李周1

作者信息

  • 1. 中国科学院长春光学精密机械与物理研究所, 长春 130033
  • 2. 中国科学院大学, 北京 100039
  • 折叠

摘要

Abstract

As is well known, the f/number of a cooled infrared detector is determined by the aperture and position of the internal cold shield. Moreover, the f/number can be changed by inserting a warm shield in front of the detector. In order to reduce the stray radiation introduced by an ordinary planar warm shield, we propose a method of varying f/number of the infrared detector based on a well-designed spherical reflecting warm shield in this paper. First, an infrared radiation model is established in order to analyze the influence of the stray radiation introduced by the ordinary planar warm shield. Then the design principle of the spherical reflecting warm shield is put forward. By changing the surface shape and emission characteristics, the stray radiation introduced by the ordinary planar warm shield can be obviously reduced. Hence it is beneficial to maintain the performance of the detector effectively while the f/number is changed. To validate the proposed method, a spherical reflecting warm shield and an ordinary planar warm shield are designed to vary the f/number of a cooled infrared detector respectively. To compare the influences of the two warm shields on the cooled infrared detector, radiometric calibration experiments are conducted in a high-low-temperature test chamber. The analyses and experimental results show that the stray radiation of spherical reflecting warm shield is far less than that of the ordinary planar warm shield. Moreover, the noise equivalent temperature difference introduced by the designed spherical reflecting warm shield is lower. Therefore it is indeed better than an ordinary planar warm shield in ensuring the performance of an infrared imaging system.

关键词

制冷型红外探测器/f数/球面反射温阑/杂散辐射

Key words

cooled infrared detector/f/number/spherical reflecting warm shield/stray radiation

引用本文复制引用

常松涛,田棋杰,何锋赟,余毅,李周..基于球面反射温阑的红外探测器变f数设计[J].物理学报,2017,66(15):52-63,12.

物理学报

OA北大核心CSCDCSTPCDSCI

1000-3290

访问量0
|
下载量0
段落导航相关论文