ABX PENTRA60全自动血细胞分析仪故障维修四例

Four Troubleshooting Cases of AbX pENTRA60 Automatic blood Cell Analyzer

张涵1

作者信息

  • 1. 马鞍山市中医院 设备科,安徽 马鞍山 243000
  • 折叠

摘要

Abstract

Four troubleshooting cases of ABX PENTRA60 automatic blood cell analyzer in our hospital were analyzed, the maintenance thoughts and methods were discussed. It was found that most of the faults were concentrated on pipeline, magnetic valve and sampling needle through researching. Faults rate could be decreased by adopting periodic maintenance, and the reason of faults could be found quickly in case that medical engineer were familiar with pipeline diagram, circuit diagram and scatter plot, which could promote targeted maintenance and guarantee clinic working.

关键词

血细胞分析仪/电磁阀/采样针/本底/负压值

Key words

blood cell analyzer/magnetic valve/sampling needle/background/negative pressure value