Abstract
Objective:To carry out the detection of quality control for defibrillator and monitor so as to ensure its safety and effectiveness in clinical practice.Methods: Impulse 7000DP defibrillator analyzer was used to detect and analyze the defibrillator and monitor, and the relevant test items included delivered energy, charging time, delay time under synchronizer and the indication deviation of heart rate.Results: In recent two years, the qualified rates including delivered energy, charging time, delay time under synchronizer and indication deviation of heart rate of defibrillator and monitors of 82 sets or times on which quality control were carried out was 100%, and then put forward the solutions and countermeasures for the problems in the testing.Conclusion: It is necessarily contents to carry out the detection of quality control for ensuring the safety and effectiveness of defibrillator and monitor in the clinical practice. To explore the regional management program of quality control will promote the development of quality control within the region.关键词
除颤监护仪/除颤器分析仪/质量控制/计量特性Key words
Defibrillator and monitor/Defibrillator analyzer/Quality control/Metrological characteristics分类