| 注册
首页|期刊导航|中国医疗设备|医疗器械非相干光视网膜蓝光 辐射危害研究

医疗器械非相干光视网膜蓝光 辐射危害研究

孟祥峰 李宁 刘艳珍 王浩 任海萍

中国医疗设备2017,Vol.32Issue(9):1-5,5.
中国医疗设备2017,Vol.32Issue(9):1-5,5.DOI:10.3969/j.issn.1674-1633.2017.09.001

医疗器械非相干光视网膜蓝光 辐射危害研究

Study on the Biohazard of Incoherent Blue Light from Medical Device on the Retina

孟祥峰 1李宁 1刘艳珍 1王浩 1任海萍1

作者信息

  • 1. 中国食品药品检定研究院 医疗器械检定所,北京 102629
  • 折叠

摘要

Abstract

Objective To determine the impact of different testing conditions on the evaluation of blue light biohazard on the retina and product classification. Methods In combination with the classification standard of light radiation hazard, the biomechanism of retinal bluelight hazard was analyzed. Blue light hazard on the retina from UV light and LED light was tested under different conditions separately (including distance, aperture stop, field stop). Results Blue light containing UV component generated more hazard on aphakic eyes (infants less than 2 years old) than regular eyes, which would probably change classification results, while lower UV proportion didn't show significant impact. The testing device didn't change the luminance of the light source. The difference of testing condition only changed the selected range of radiant angle and area with different results. Conclusion The difference of testing conditions has impact on radiant exposure, but in general, it doesn't change classification results. To assure the accuracy of radiant exposure measurement, appropriate experimental condition should be selected based on the actual circumstance of light sources following the most unfavorable rule.

关键词

医疗器械/非相干光/视网膜/蓝光危害/无晶状体眼/对向角

Key words

medical device/incoherent light/retina/blue light hazard/aphakic eye/angular subtense

分类

医药卫生

引用本文复制引用

孟祥峰,李宁,刘艳珍,王浩,任海萍..医疗器械非相干光视网膜蓝光 辐射危害研究[J].中国医疗设备,2017,32(9):1-5,5.

基金项目

中国食品药品检定研究院中青年基金课题"医用LED设备光辐射危害评价与检测方法研究"(2015C01) (2015C01)

人社部留学人员科技活动项目择优资助课题"医疗器械光辐射安全评价研究"资助. ()