中国医疗设备2017,Vol.32Issue(10):102-104,3.DOI:10.3969/j.issn.1674-1633.2017.10.028
Philips EPIQ 7彩色超声诊断成像仪故障维修两例
Two Cases of Troubleshooting for Philips EPIQ 7 Color Ultrasound Diagnostic Imaging System
杨新星 1朱霆 1巩雪 1孙园园 1于铭 1刘丽文1
作者信息
- 1. 第四军医大学西京医院 超声科,陕西 西安 710032
- 折叠
摘要
Abstract
This paper mainly analyzed two special troubleshooting cases of Philips EPIQ 7 color ultrasound diagnostic imaging system, and the maintenance method was discussed. Philips EPIQ 7 in the ultrasound department of our hospital had peculiar smell, this paper analyzed causes of the smell, found source of the trouble and repaired it. Results showed that bad contact of operating tabs on the instrument panel result in burning odor, and small amount of ash accumulation result in error report. This two fault cases analyzed in this paper are rare, so it need patience to find fault source. Regular maintenance can decrease faults occurrence rate, and ensure normal using of the machine.关键词
彩色多普勒/超声检查/操作按键/设备维护/仪器面板Key words
color doppler/ultrasonic inspection/operation keys/equipment maintenance/instrument panel分类
医药卫生引用本文复制引用
杨新星,朱霆,巩雪,孙园园,于铭,刘丽文..Philips EPIQ 7彩色超声诊断成像仪故障维修两例[J].中国医疗设备,2017,32(10):102-104,3.