中国医疗设备2017,Vol.32Issue(10):105-109,5.DOI:10.3969/j.issn.1674-1633.2017.10.029
电阻抗断层成像中抑制电极移动伪影方法的研究进展
Research Status Regarding to Inhibiting Electrode Perturbation Caused Artefacts in Electrical Impedance Tomography
摘要
Abstract
Electrical impedance tomography imaging has been widely applied in medical field, such as lung function, early detection of breast cance, brain function imaging and so on. In the process of imaging, the variation of the electrode on the boundary of the measuring object will cause the error of model parameter in the electrical impedance imaging algorithm, resulting in a large number of artifacts in the reconstructed image. To solve this problem, firstly, this paper summarized several algorithms to improve the noise interference, and gave a brief introduction to the physical basis of these algorithms. Then, the principle and characteristics of the current several anti-electrode position change interference were summarized. At the end of the paper, the algorithms were analyzed and the future research and application of those algorithms were prospected.关键词
电阻抗成像/电极位置变动/算法/图像伪影/差分近似法/有限元几何特征法Key words
electrical impedance tomography/electrode boundary perturbations/algorithm/image artifacts/difference approximation/finite element geometry method分类
医药卫生引用本文复制引用
陈荣庆,徐灿华,董秀珍,付峰..电阻抗断层成像中抑制电极移动伪影方法的研究进展[J].中国医疗设备,2017,32(10):105-109,5.基金项目
国家自然科学基金面上项目(51477176). (51477176)