临床口腔医学杂志2017,Vol.33Issue(10):585-588,4.DOI:10.3969/j.issn.1003-1634.2017.10.003
羟基磷灰石种植体涂层体内消减的相关研究
The analysis of the degradation of HA coating of a loaded implant
摘要
Abstract
Objective:To analyze the degradation of hydroxyapatite(HA)coating of a failed implant after long term function. Methods:The HA-coated implant functioning for 23 years was extracted due to peri-implantitis. The surface was an-alyzed by gross observation,scanning electron microscopy( SEM) for morphology,and energy dispersive spectrometer( EDS) for elements analysis respectively. Results:By gross observation and SEM test,no obvious coating-like structure was found,ex-cept for some micro-convexes. EDS analysis demonstrated that calcium and phosphorus were detected(14. 27%,9. 15%) at the convexes. For most of the areas,no or trace calcium and phosphorus(1. 17%,0. 82%) was found. Conclusion:The spe-cific case analysis showed that HA coating of the implant almost degraded after the long term function in the human body.关键词
羟基磷灰石/牙种植体/扫描电子显微镜/X射线能谱分析Key words
Hydroxyapatite/Dental Implant/Scanning electron microscopy/Energy dispersive spectrometer分类
医药卫生引用本文复制引用
程孟文,周毅,王贻宁,陈琦..羟基磷灰石种植体涂层体内消减的相关研究[J].临床口腔医学杂志,2017,33(10):585-588,4.基金项目
国家自然科学基金(81200812) (81200812)
武汉市应用基础研究计划项目(2016060101010043) (2016060101010043)